A practical reference on HPLC: what it is, how it behaves, what the literature reports, and where the honest uncertainties sit.
This page was last updated on 2026-01-05 and is reviewed periodically as new material appears.
Separation performance depends on particle size, pore size, column length, and the chemistry of the stationary phase. Smaller particles generally improve efficiency but require higher pressure and suitable instrumentation. The mobile phase often contains buffers and organic solvents that influence retention and selectivity. Testing labs select conditions based on the analytes, sample matrix, and required sensitivity. Method development frequently involves screening several columns and solvent mixtures before a final set of conditions is chosen.
High-performance liquid chromatography is an analytical technique that separates components in a liquid sample by passing them through a packed column under pressure. A pump delivers a mobile phase at a controlled flow rate, and an injector introduces the sample into the stream. Differences in how analytes partition between the mobile phase and the stationary phase cause them to exit the column at different times. Detection then records a signal proportional to the amount of each separated substance. The resulting chromatogram provides retention times and peak areas for identification and quantification.
System suitability testing is performed before and during analytical runs to confirm that the instrument and method are working as expected. Common checks include retention time, peak area, resolution between critical pairs, tailing factor, and theoretical plate count. Results are compared with predefined limits, and a failed check requires investigation before sample results are reported. Quality control samples at low, middle, and high concentrations are injected at intervals to monitor accuracy and precision. Blank injections detect carryover and contamination, while control charts track performance over time.
Data handling and documentation are central to HPLC quality control. Electronic systems should have audit trails that record changes to methods, sequences, and results. Integration parameters, such as peak baseline and threshold, can affect reported areas and must be defined in advance. Out-of-specification results trigger a structured investigation that may include reanalysis, instrument checks, and review of sample preparation. Regulatory inspections often examine raw data, audit trails, and training records to verify that reported results are traceable and reliable.
| Property | Value | Notes |
|---|---|---|
| Separation principle | Differential partitioning | Analytes distribute between mobile and stationary phases. |
| Mobile phase | Liquid solvent mixture | Composition controls retention and selectivity. |
| Stationary phase | Packed column particles | Often chemically bonded silica. |
| Typical detector | UV-Vis or photodiode array | Mass spectrometry is also common. |
| Common synonym | High-performance liquid chromatography | Abbreviated as HPLC. |
Quality control samples are inserted at intervals to monitor accuracy and precision throughout a batch. Blank samples detect contamination, while spiked samples assess recovery from the sample matrix. Calibration standards establish the relationship between detector response and concentration, and control samples are prepared independently from them whenever possible. Laboratories also participate in proficiency testing and maintain audit trails, instrument logs, and reagent records. Ongoing review of control charts can reveal trends before they cause out-of-specification results.
Method validation demonstrates that an HPLC procedure is suitable for its intended purpose. Common validation parameters include accuracy, precision, specificity, linearity, range, limit of detection, limit of quantification, and robustness. Accuracy reflects agreement with a reference value, while precision describes repeatability under defined conditions. Specificity shows whether the method can measure the analyte in the presence of impurities or matrix components. Validation documents are reviewed before a method is used for routine testing or regulatory submissions.
Separation in HPLC depends on the chemistry of the stationary phase, the composition of the mobile phase, and the physical properties of the column. Reverse-phase separations use a nonpolar stationary phase and a polar mobile phase, and they are common for many organic compounds. Ion-exchange, size-exclusion, and normal-phase modes serve other classes of analytes. Gradient elution changes solvent strength over time, while isocratic elution holds it constant. Flow rate, temperature, particle size, and column length all influence peak shape and resolution. Detection may use ultraviolet absorbance, fluorescence, refractive index, or mass spectrometry, depending on the analyte and the required sensitivity.
Routine HPLC testing compares a sample result with a calibration curve prepared from known reference standards. Peak area or peak height is plotted against concentration, and the curve is used to estimate unknown amounts. Retention time supports tentative identification when compared with a standard, though mass spectrometry or another confirmatory method may be needed for definitive identification. Pre-run checks verify repeatability, resolution, and peak symmetry before sample analysis. Limits of detection and quantification describe the smallest amounts that can be reliably observed or measured. Sample preparation, filtration, and degassing help prevent column damage and inconsistent results.
Detection commonly uses ultraviolet-visible absorbance, fluorescence, refractive index, or mass spectrometry. Ultraviolet detection depends on molecular chromophores that absorb light at specific wavelengths. Mass spectrometry provides mass information and sensitive quantification, often after electrospray ionization. Before sample batches, performance checks examine resolution, elution time repeatability, peak symmetry, and plate count. Matrix effects and co-elution remain recognized uncertainties; formal validation studies and orthogonal detection help address them. Detector choice depends on analyte properties and required sensitivity.
High-performance liquid chromatography, or HPLC, separates dissolved compounds by passing a liquid mobile phase through a packed column. Components distribute differently between the stationary phase and the moving liquid, so they travel at different speeds and exit at different times. A detector records these eluting bands as peaks, and peak area or height relates to amount. The technique supports testing in pharmaceuticals, foods, environmental samples, and industrial chemicals. Quantification usually depends on calibration with known standards.
Method validation examines whether an HPLC procedure is suitable for its intended purpose. Common parameters include accuracy, precision, specificity, linearity, range, detection limit, quantification limit, and robustness. Accuracy describes closeness to a true or accepted value, while precision describes agreement among repeated measurements. Specificity shows whether the method can measure the analyte without interference from related substances. Robustness tests small deliberate changes in flow, temperature, or solvent composition. Validation is not a one-time event; methods may need partial revalidation after changes to instruments, columns, sample handling, or specification limits. Regulatory guidance provides frameworks, but some details remain method-specific.
Regulatory and pharmacopeial texts shape how HPLC testing is performed and documented. The International Council for Harmonisation provides validation guidance, while pharmacopeias publish general chromatography chapters and monographs for specific materials. Accreditation standards such as ISO/IEC 17025 address laboratory competence and traceability. Inspectors may review instrument qualification, analyst training, reference material control, and electronic records. Open questions include how best to validate methods for new complex products and how to handle automated data processing. Laboratories generally resolve these issues through risk assessment, method lifecycle management, and documented scientific justification.
Coomassie Blue is the most commonly used non-covalent stain in SDS polyacrylamide gel electrophoresis for protein quantification. The staining dye binds to the protein bands and creates a blue color that can be detected visually. Coomassie Brilliant Blue R-250 (red), is typically used for electrophoresis, while Coomassie Brilliant Blue G-250 (green), for Bradford Assay. The limitation of this dye is that it is non-specific, and will bind to almost any protein in solution, and is less sensitive. Another common method of visualization of proteins in the gel is silver staining, where soluble silver ions permanently mark proteins and are reduced by formaldehyde to form a brown precipitate. Silver staining is a more sensitive staining method when compared to Coomassie Blue, however, results are more vulnerable to contamination.
A flagellate is a cell or organism with one or more whip-like appendages called flagella. The word flagellate also describes a particular construction (or level of organization) characteristic of many prokaryotes and eukaryotes and their means of motion. The term presently does not imply any specific relationship or classification of the organisms that possess flagella. However, several derivations of the term "flagellate" (such as "dinoflagellate" and "choanoflagellate") are more formally characterized.
The terms "active" and "passive" are simple but important terms in the world of automotive safety. "Active safety" is used to refer to technology assisting in the prevention of a crash and "passive safety" to components of the vehicle (primarily airbags, seatbelts and the physical structure of the vehicle) that help to protect occupants during a crash. Crash avoidance systems and devices help the driver — and, increasingly, help the vehicle itself — to avoid a collision. This category includes: The vehicle's headlamps, reflectors, and other lights and signals The vehicle's mirrors The vehicle's brakes, steering, and suspension systems A subset of crash avoidance is driver assistance systems, which help the driver to detect obstacles and to control the vehicle. Driver assistance systems include:
Ketone bodies can be used as fuels, yielding 22 ATP and 2 GTP molecules per acetoacetate molecule when oxidized in the mitochondria. Ketone bodies are transported from the liver to other tissues, where acetoacetate and beta-hydroxybutyrate can be reconverted to acetyl-CoA to produce reducing equivalents (NADH and FADH2), via the citric acid cycle. Ketone bodies cannot be used as fuel by the liver, because the liver lacks the enzyme β-ketoacyl-CoA transferase, also called thiolase. Acetoacetate in low concentrations is taken up by the liver and undergoes detoxification through the methylglyoxal pathway which ends with lactate. Acetoacetate in high concentrations is absorbed by cells other than those in the liver and enters a different pathway via 1,2-propanediol. Though the pathway follows a different series of steps requiring ATP, 1,2-propanediol can be turned into pyruvate.
Sources: en.wikipedia.org
Brunton Medal 1979, Sir Francis Avery Jones research medal of the British Society of Gastroenterology, the highest research award that the national gastroenterology body confers on its researchers. Fellowships: Royal Colleges of Physicians of Glasgow (1975) London (1977) Edinburgh (1981) Royal College of Pathologists (1984) Fellow of the Royal Society of Edinburgh (1990) The Anne Ferguson building at the Western General Hospital, Edinburgh, is named after her. In 1966, she married John Ferguson, a sociology lecturer at the University of Strathclyde, and together they adopted two children, a girl and a boy. John Ferguson died of cancer in 1989. She married Professor Gerald Collee, emeritus Professor of Medical Microbiology at the University of Edinburgh in 1995. In her youth, Ferguson represented the University of Glasgow in athletics as a middle-distance runner. She was also on the Scottish women's basketball team. She enjoyed hill walking and mountaineering, and once visited the Himalayas with her first husband. Ferguson died of pancreatic cancer in Edinburgh on 21 December 1998.
Gs exerts its effects via two pathways. Firstly, it directly opens L-type calcium channels (LTCC) in the plasma membrane. Secondly, it renders adenylate cyclase activated, resulting in an increase of cAMP, activating protein kinase A (PKA) which in turn phosphorylates several targets, such as phospholamban, LTCC, Troponin I (TnI), and potassium channels. The phosphorylation of phospholamban deactivates its own function which normally inhibits SERCA on the sarcoplasmic reticulum (SR) in cardiac myocytes. Due to this, more calcium enters the SR and is therefore available for the next contraction. LTCC phosphorylation increases its open probability and therefore allows more calcium to enter the myocyte upon cell depolarisation. Both of these mechanisms increase the available calcium for contraction and therefore increase inotropy. Conversely, TnI phosphorylation results in its facilitated dissociation of calcium from troponin C (TnC) which speeds the muscle relaxation (positive lusitropy). Potassium channel phosphorylation increases its open probability which results in shorter refractory period (because the cell repolarises faster), also increasing lusitropy. Furthermore, in nodal cells such as in the SA node, cAMP directly binds to and opens the HCN channels, increasing their open probability, which increases chronotropy.
The ENCODE project proposes to map all of the DHSs in the human genome with the intention of cataloging human regulatory DNA. DHSs mark transcriptionally active regions of the genome, where there will be cellular selectivity. So, they used 125 different human cell types. This way, using the massive sequencing technique, they obtained the DHSs profiles of every cellular type. Through an analysis of the data, they identified almost 2.9 million distinct DHSs. 34% were specific to each cell type, and only a small minority (3,692) were detected in all cell types. Also, it was confirmed that only 5% of DHSs were found in TSS (Transcriptional Start Site) regions. The remaining 95% represented distal DHSs, divided in a uniform way between intronic and intergenic regions. The data gives an idea of the great complexity regulating the genetic expression in the human genome and the quantity of elements that control this regulation. The high-resolution mapping of DHSs in the model plant Arabidopsis thaliana has been reported. Total 38,290 and 41,193 DHSs in leaf and flower tissues have been identified, respectively.
$40 million for the COVID-19 Genomics Network $23 million for the Vaccine and Infectious Disease Organization-International Vaccine Centre $29 million for the National Research Council of Canada $600 million through the Strategic Innovation Fund $10.3 million over 10 years, $5 million bonus to support the Canadian Immunization Research Network $114.9 million through the Canadian Institutes of Health Research A Vaccine Surveillance Reference Group (VSRG) was also established within the CITF to monitor the safety and effectiveness of COVID-19 vaccines made available in Canada. The task force was to also estimate how many Canadians were immune to the SARS-CoV-2 virus. Catherine Hankins reported that less than 1% of 10,000 samples tested positive for antibodies to SARS-CoV-2. The CITF also found that one in 100 Canadians were infected with COVID-19 during the first wave of the pandemic. Canadian Blood Services analyzed over 30,000 blood samples in total. The CITF Board was composed of doctors, infectious disease experts, and policy makers.
Aaron R. Wheeler is a Canadian chemist who is a professor of chemistry and biomedical engineering at the University of Toronto since 2005 with cross-appointment at Institute of Biomedical Engineering and Terrence Donnelly Centre for Cellular and Biomolecular Research. His academic laboratory is located at Lash Miller Chemical Laboratories and Terrence Donnelly Centre for Cellular and Biomolecular Research at the University of Toronto. In 2005, Wheeler was appointed as assistant professor and Tier II Canada Research Chair then promoted to associate professor in 2010, full professor in 2013, and in 2018 he became the Tier I Canada Research Chair in Microfluidic Bioanalysis. Wheeler did his undergraduate studies at Furman University in Greenville, SC then he joined Stanford University from 1997 to 2003 to obtain his Ph.D. in chemistry under supervision of Richard Zare . Following graduation, he took a two-year NIH postdoctoral fellowship at UCLA till 2005.
Sources: en.wikipedia.org
HPLC separates and detects individual compounds in a liquid sample, producing peaks at characteristic retention times. Peak area or height can be used to estimate concentration when calibrated with known standards. It does not identify unknown compounds with certainty unless additional detectors or reference materials are used.
Pressure drives the liquid mobile phase through a column packed with small particles. Without pressure, flow would be very slow or stop because the packed bed resists liquid movement. Modern pumps maintain a steady flow despite the resistance.
A chromatogram is a plot of detector signal against time after sample injection. Each peak represents a compound or group of compounds eluting from the column. Retention time and peak area are the main measurements read from the plot.
System suitability is a set of checks that confirm the instrument and method perform within limits before sample analysis. It typically includes resolution, tailing factor, retention time, and peak area reproducibility. If a check fails, the run is invalidated until the cause is resolved.